dc.description.abstract This study introduced a shape memory alloy (SMA)-based smart knee spacer for total knee arthroplasty (TKA). Subsequently, a 3D CAD model of a smart tibial component of TKA was designed in Solidworks software, and verified using a finite element analysis in ANSYS Workbench. The two major properties of the SMA (NiTi), the pseudoelasticity (PE) and shape memory effect (SME), were exploited, modelled, and analysed for a TKA application. The effectiveness of the proposed model was verified in ANSYS Workbench through the finite element analysis (FEA) of the maximum deformation and equivalent (von Mises) stress distribution. The proposed model was also compared with a polymethylmethacrylate (PMMA)-based spacer for the upper portion of the tibial component for three subjects with body mass index (BMI) of 23.88, 31.09, and 38.39. The proposed SMA -based smart knee spacer contained 96.66978% less deformation with a standard deviation of 0.01738 than that of the corresponding PMMA based counterpart for the same load and flexion angle. Based on the maximum deformation analysis, the PMMA-based spacer had 30 times more permanent deformation than that of the proposed SMA-based spacer for the same load and flexion angle. The SME property of the lower portion of the tibial component for fixation of the spacer at its position was verified by an FEA in ANSYS. Wherein, a strain life-based fatigue analysis was performed and tested for the PE and SME built spacers through the FEA. Therefore, the SMA-based smart knee spacer eliminated the drawbacks of the PMMA-based spacer, including spacer fracture, loosening, dislocation, tilting or translation, and knee subluxation.
dc.identifier.citation Medical Engineering and Physics. v.55
dc.identifier.issn 13504533
dc.identifier.uri 10.1016/j.medengphy.2018.03.001
dc.identifier.uri https://www.sciencedirect.com/science/article/abs/pii/S1350453318300432
dc.identifier.uri https://dspace.uohyd.ac.in/handle/1/6244
dc.subject CAD
dc.subject Finite element analysis
dc.subject Knee kinetics
dc.subject Knee osteoarthritis
dc.subject PMMA
dc.subject SMA
dc.subject TKA
dc.title Shape-memory-alloy-based smart knee spacer for total knee arthroplasty: 3D CAD modelling and a computational study