Factors influencing the sensitivity of the CDMPI sensor

dc.contributor.author Kollu, Pratap
dc.contributor.author Yoon, Seok Soo
dc.contributor.author Kim, Chong Oh
dc.contributor.author Kim, Cheol Gi
dc.date.accessioned 2022-03-27T06:45:30Z
dc.date.available 2022-03-27T06:45:30Z
dc.date.issued 2007-01-01
dc.description.abstract A highly sensitive magnetic sensor using the GMI (Giant Magneto Impedance) effect has been developed. The sensor performance is studied and estimated. The sensor circuitry consists of a Square wave generator (driving source), a sensing element in a form of composite wire of a 20 μm copper core electrodeposited with a thin layer of son magnetic material (Ni80Fe20), two amplifier stages for improving the gain from the sensor element., switching mechanism, scaler circuit, an ac power source driving the permeability of the magnetic coating layer of the sensing element into a dynamic state, and a signal pickup LC circuit formed by a pickup coil and an capacitor. The results showed that for high sensitivity and resolution (10-7 T), the frequency and magnitude of the ac driving current through the sensing element each has an optimum value, the resonance frequency of the signal pickup LC circuit should be equal to or twice as the driving frequency on the sensing element, and the anisotropy of the magnetic coating layer of the sensing wire element should be longitudinal.
dc.identifier.citation Solid State Phenomena. v.124-126(PART 1)
dc.identifier.issn 10120394
dc.identifier.uri 10.4028/3-908451-31-0.887
dc.identifier.uri http://www.crossref.org/deleted_DOI.html
dc.identifier.uri https://dspace.uohyd.ac.in/handle/1/10080
dc.subject CDMPI
dc.subject Giant magnetoimpedance
dc.subject Permalloy
dc.subject Sensor
dc.title Factors influencing the sensitivity of the CDMPI sensor